Category:

SL 176

True Double Beam Atomic Absorption Spectrophotometer

Flame Atomic Absorbance Spectrophotometer (AAS), Cost effective & easy-to-use solutions for efficient and accurate trace elemental analysis. Simple, Integral gas control system with panel flow meters.

Description
ELICO's SL- 176, Atomic Absorption Spectrophotometer comes with semi-automated gas control system and delivers a new level of flexibility and simplicity. It can work as standalone mode for routine analysis and can also be PC Controlled.
Applicable Industries
Features
Burner Alignment
Double Beam
LCD
Auto Ignition
PC
  • Microprocesor and Microcontroller based user friendly, Menu driven with printer interface and optional PC attachment through USB Interface
  • High Energy True Double Beam Optics with Narrow Beam, fully sealed and Quartz over coated mirrors
  • 6/8 lamp turret with independent power supplies for warm-up
  • Background correction with D2 lamp source
  • Integral gas control system with panel flow meters
  • Precise Knobs for Vertical, Horizontal and Rotational alignment of Titanium Burner
  • Modes of operation include Absorbance, Emission, Concentration and Standard Addition
  • Optional : 21 CFR Part 11 PC Software, Auto Sampler, Hydride and Cold Vapour Generator
Specifications
SPECTRAL
Range 185 to 900 nm
Bandwidth Automatic variable slit from 0 to 2.5 nm continuously variable with increment of 0.1 nm
Readability 0.1 nm
Accuracy ± 0.5 nm
Repeatability ± 0.2 nm
PHOTOMETRIC
Range 0 to 2 Abs
Accuracy ± 0.010 Abs at 1.0 Abs
Integration Time 0.04 to 99 sec
LIGHT SOURCE
6/8 Lamps Automatic Turret with Independent Power Supply compatible to coded and non coded HC Lamps.
MONOCHROMATOR
Czerney Turner 1/3 m with 1800 lines/mm Holographic Grating and Hard Coded Optics.
BACK GROUND CORRECTION
Deuterium Lamp (D2)
BURNER
Titanium construction, Premix Laminar Flow100mm burner for Air – Acetylene (C2H2)
Optional – 50 mm burner for Nitrous Oxide (N2O) –  Acetylene (C2H2)
The Spray Chamber is made of Inert Material .i.e., PTFE (Teflon) fitted with adjustable impact bead Aerosol
NEUBULIZER
Made of fully Inert Material with 90% Platinum and 10% Iridium Alloy Capillary
SAMPLE FEED RATE
3 to 8 ml/min (Adjustable)
DETECTOR
Wide Range Photo Multiplier Tube (PMT)
FLOW CONTROL
Integral Gas Control System with Panel Flow Meters
SYSTEM UPGRADATION
Instrument can be upgraded to Auto Sampler, Hydride and Cold vapour Generator
ACCESSORIES SUPPLIED
With Instrument 100 mm burner for Air – C2H2 , Nebulizer with SS Capillary, Back Ground Correction with D2 Lamp, Oil free Air Compressor with 100% Dustfree Air Filter, Acetylene (C2H2) regulator with Flashback Arrester.
At extra Cost 50 mm burner for N2O – C2H2 , Nebulizer with Pt – Ir Capillary, N2O Regulator, Nitrogen Regulator, HC Lamps for desired Elements coded or non coded, Hydride and Cold vapour Generator, Printer, Computer, PC Software, Exhaust Fumehood Chamber, Servo Controlled Voltage Stabilizer of suitable rating, Acetylene Gas Cylinder, Nitrous Oxide Gas Cylinder and Nitrogen Gas Cylinder.

Reviews

There are no reviews yet.

Be the first to review “SL 176”

Your email address will not be published. Required fields are marked *

Start typing and press Enter to search

Shopping Cart

No products in the cart.